smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps

Recall: Z-0466-2025 · Initiated June 26, 2024

Recall

Recall Number
Z-0466-2025
Event Number
95483
Firm
Smiths Medical ASD, Inc.
FEI Number
3012307300
Product Code
MEA
Status
Open, Classified
Root Cause
Device Design
Initiated
June 26, 2024
Posted
November 15, 2024
Address
6000 Nathan Ln N, Minneapolis, MN, 55442-1690

Description

smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps

Reason

The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.

Action

Smiths Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 06/26/2024 via letter distributed by a traceable means. The notice explained the issue, potential risk, and requested the following: Locate all Model 21-0270-25 AC Adapters in your possession. Inspect the AC Adapter for damage to the input plug, missing AC prongs, or exposed metal contacts. If damaged, remove the AC Adapter from service and submit a request for credit to Smiths Medical at 800-258-5361. Distributors were directed to forward to the notice to those to whom the they distributed the device with the request to complete the response form and return it to [email protected].

Distribution

Worldwide distribution.

Quantity

156,400 AC adapters