SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. SoftMic is a laboratory information system to be used in medical research, clinical, and public health laboratories by knowledgeable, trained, and experienced personnel. SoftMic enables the user to handle data processing and laboratory activities for microbiology and serology testing, for the purpose of storing, managing, querying, and reporting laboratory data.
Recall
- Recall Number
- Z-0466-2012
- Event Number
- 60459
- Firm
- SCC Soft Computer
- FEI Number
- 1000307651
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 25, 2011
- Posted
- January 11, 2012
- Terminated
- October 3, 2014
- Address
- 5400 Tech Data Drive, Clearwater, FL, 33760
Description
SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. SoftMic is a laboratory information system to be used in medical research, clinical, and public health laboratories by knowledgeable, trained, and experienced personnel. SoftMic enables the user to handle data processing and laboratory activities for microbiology and serology testing, for the purpose of storing, managing, querying, and reporting laboratory data.
On 08/25/2011, SCC Soft Computer initiated a correction on the following SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. For clients using SoftMic Result Entry, when a certain series of keystrokes on the test grid are performed while entering results and statusing a test, the test status may not be maintained upon saving. The status may be removed or copied from another test on the order
SCC Soft Computer initiated a Field Correction dated August 25, 2011. to all affected customers. Customers with the affected software have been notified and are scheduling receipt of the hot fix.. Further questions call (727) 789-0100.
Worldwide Distribution -- USA (nationwide) and Canada.