FDA Recall Terminated

Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 3 Model Numbers are used for this same Burette Administration Set described as 110 inch length/279 cm, 16.5ml priming volume, 20 drops/ml, 3 standard injection sites. Model Number I8C1230J (Reorder No. 21-0301-25) for product distributed to Japan. Model Number I8C1230 (Reorder No. (21-0318-25) for product distributed Internationally. Model Number 8C1230 (Reorder No. 21-0305-01) for product distributed within the United States.

Recall: Z-0466-05 · Initiated December 3, 2004

Recall

Recall Number
Z-0466-05
Event Number
30603
Firm
Smiths Medical MD, Inc.
FEI Number
2183502
Product Code
FPA
Status
Terminated
Root Cause
Other
Initiated
December 3, 2004
Posted
February 2, 2005
Terminated
June 25, 2006
Address
1265 Grey Fox Rd, Saint Paul, MN, 55112-6929

Description

Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 3 Model Numbers are used for this same Burette Administration Set described as 110 inch length/279 cm, 16.5ml priming volume, 20 drops/ml, 3 standard injection sites. Model Number I8C1230J (Reorder No. 21-0301-25) for product distributed to Japan. Model Number I8C1230 (Reorder No. (21-0318-25) for product distributed Internationally. Model Number 8C1230 (Reorder No. 21-0305-01) for product distributed within the United States.

Reason

Administration Sets which are used with selected infusion pump systems may have a defect within the fluid path above the upper chamber of the set cassette which does not allow the pump valve to close completely and can allow fluid to leask past and may cause an inaccurate infusion rate..

Action

Customers were sent a product Safety & Recall notification via email or regular mail begining 12/03/04 through 12/08/04. A Product Recall form requested the customer for feedback as to Quantity Used, Quantity to be Retained and Quantity to be returned . The Customer Recall Notication states that if no Administration Sets from alternative lots are available and the use of the volumetric pump is essential, follow the instructions in the attached Customer Information Bulletin, which describes how to identify a set which does not have the defect. Customers who find product subject to the recall are to contact Smith Medical Customer Service Department to make arrangements to have the affected products returned to Smiths.

Distribution

worldwide.

Quantity

31,320 units total: 6,150 units (205 cases) to Japan, 17,970 units (599 cases) Internationally &7,200 units (240 cases) within USA. ea case 30 units.