FDA Recall Terminated

Depuy Preservation Uni-Compartmental Knee Fixed Tibial Tray, cemented, Right Medial/Left Lateral, Size 4, sterile; Part number 1498-23-004

Recall: Z-0463-2007 · Initiated November 15, 2006

Recall

Recall Number
Z-0463-2007
Event Number
36849
Firm
Depuy Orthopaedics, Inc.
FEI Number
1818910
Product Code
HRY
Status
Terminated
Root Cause
Other
Initiated
November 15, 2006
Posted
January 11, 2007
Terminated
April 30, 2008
Address
700 Orthopaedic Dr, Warsaw, IN, 46582-3994

Description

Depuy Preservation Uni-Compartmental Knee Fixed Tibial Tray, cemented, Right Medial/Left Lateral, Size 4, sterile; Part number 1498-23-004

Reason

Chamfer Step Defect-The surgeon may have difficulty in assembling /seating the polyethylene insert into the tray due to a Chamfer step defect.

Action

The firm's sales force was notified of the recall via email sent 11/15/06. The sales force was instructed to deliver a recall letter to personnel in those facilities that had purchased the products, and the recall letters instruct the facilities to quarantine the product and respond to Depuy concerning product returns. The sales force was provided with a Dear Doctor letter for issuance to implanting surgeons at the sales forces' discretion. The recall was extended to Sizes 1-5 on Dec 5, 2006 with a second recall letter.

Distribution

Class II Recall - Worldwide Distribution ---USA and countries of Argentina and Colombia.

Quantity

558 for Recalls Z-0456-0465-2007