FDA Recall Terminated

Philips Medical Systems, GEMINI GXL 6, 10, & 16 slice PET/CT Systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems, Philips Medical Systems, Cleveland, OH 44143

Recall: Z-0462-2008 · Initiated September 21, 2007

Recall

Recall Number
Z-0462-2008
Event Number
44934
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
KPS
Status
Terminated
Root Cause
Software design
Initiated
September 21, 2007
Posted
January 31, 2008
Terminated
March 2, 2012
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

Philips Medical Systems, GEMINI GXL 6, 10, & 16 slice PET/CT Systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems, Philips Medical Systems, Cleveland, OH 44143

Reason

Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.

Action

Philips Medical notified domestic customers via Product Safety Notification dated May 30, 2007. The product safety notification includes a Technical Bulletin notifying customers of the issue and how to avoid it. This information will be used until the software correction becomes available. The software correction will be released in October 2008. Once software is available, the firm will begin correcting the field.

Distribution

Worldwide Distribution - USA, Australia, Belgium, Canada, China, Denmark, England, Finland, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Jordan, Korea, Netherlands, Poland, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Taiwan, Turkey, and Venezuela.

Quantity

126