FDA Recall Terminated

Reagent Management System (RMS) Dimension RxL Max w/o HM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system to process samples without interruption.

Recall: Z-0460-2016 · Initiated November 10, 2015

Recall

Recall Number
Z-0460-2016
Event Number
72738
Firm
Siemens Healthcare Diagnostics Inc.
FEI Number
3002755490
Product Code
JJE
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 10, 2015
Posted
December 22, 2015
Terminated
August 23, 2017
Address
101 Silvermine Rd, Brookfield, CT, 06804-2047

Description

Reagent Management System (RMS) Dimension RxL Max w/o HM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system to process samples without interruption.

Reason

Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.

Action

Siemens sent an Urgent Medical Device Correction (UMDC # DI-16-01.A.US) and Urgent Field Safety Notice (UFSN # DI-16-01.A.OUS) letters dated November 2015 to all affected customers. The letters identified the affected product, problem and actions to be taken. For questions contact your local Siemens technical support representative.

Distribution

US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.

Quantity

3