FDA Recall Terminated

ARTISTE and ONCOR linear accelerators with G31 mother board. ARTISTE and ONCOR are manufactured by Siemens Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany. ARTISTE, ONCOR and PRIMUS family of linear accelerators deliver X-ray photon and electron radiation for therapeutic treatment of cancer.

Recall: Z-0459-2012 · Initiated November 9, 2011

Recall

Recall Number
Z-0459-2012
Event Number
60525
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2937457
Product Code
IYE
Status
Terminated
Root Cause
Component design/selection
Initiated
November 9, 2011
Posted
January 11, 2012
Terminated
April 15, 2013
Address
4040 Nelson Ave, Concord, CA, 94520-1200

Description

ARTISTE and ONCOR linear accelerators with G31 mother board. ARTISTE and ONCOR are manufactured by Siemens Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany. ARTISTE, ONCOR and PRIMUS family of linear accelerators deliver X-ray photon and electron radiation for therapeutic treatment of cancer.

Reason

Reports of gantry rotation without receiving a movement command caused by an overload within the drive signal power line.

Action

Siemens Medical Solutions USA, Inc. sent a Customer Information letter dated November 9, 2011, by federal express. The letter identified the product the problem and the action needed to be taken by the customer. The HW Update included information regarding the potential problem and provide instructions in the Safety update. Further questions call (925) 602-8157.

Distribution

Worldwide Distribution-- USA ( nationwide ) including the states of : MO, NC, OH, and OK and the countries of Turkey, Japan, Germany and Portugal

Quantity

13 active devices