The Dimension TSHL Flex reagent cartridge (Thyroid Stimulating Hormone) manufactured by Siemens Healthcare Diagnostics, Inc., Newark, DE.
Recall
- Recall Number
- Z-0459-2011
- Event Number
- 55604
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- JLW
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 12, 2010
- Posted
- November 24, 2010
- Terminated
- March 19, 2013
- Address
- 500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101
Description
The Dimension TSHL Flex reagent cartridge (Thyroid Stimulating Hormone) manufactured by Siemens Healthcare Diagnostics, Inc., Newark, DE.
Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.
Siemens issued an "Urgent Field Safety Notice" dated April 2010, informing users of the problem and instructing them to inspect cartridges for leaking fluid. Cartridges with leaking fluid should be reported to the Siemens Technical Solutions Center. The Technical Solutions Center can be contacted at 800 441 9250.
Worldwide Distribution: USA, Europe, Australia, China, Japan, Korea, and Saudi Arabia.
4187