FDA Recall Open, Classified

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001F0

Recall: Z-0458-2025 · Initiated October 3, 2024

Recall

Recall Number
Z-0458-2025
Event Number
95498
Firm
Getinge Usa Sales Inc
FEI Number
3013876692
Product Code
FQO
Status
Open, Classified
Root Cause
Component design/selection
Initiated
October 3, 2024
Posted
November 14, 2024
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035

Description

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001F0

Reason

Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

Action

Maquet GmbH, a subsidiary of Getinge (Maquet/Getinge) issued Urgent Medical Device Correction (FSCA 8010652-09/25/2024-002-C) letter on 03 October 2024, via FedEx delivery. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the MEERA Operating Tables with the Serial numbers listed in this notice. Note: The Product REF number (REF) and the Serial Number (SN) can be found on the product label. If the error code "50037" appears-(note below picture of the remote control displaying the Error Code), it will be displayed on the IR-Hand Control. In such cases, you can: a) Restart the IR-Hand Control by releasing the key and pressing the key again, b) Switch to the corded control device, which is unaffected by this issue, or c) Use the override control panel on the operating table. The table remains safe to use as described in the instructions for use (IFU700001XYEN05_01 / IFU710001XYEN07_01 / IFU720001XYEN14_01), with special focus on: " Chapter 7.3: Explains error messages and recovery steps. " Chapter 4.3: Details the use of the override control panel. The override control panel provides immediate access to table movements, ensuring continued operability even in the unlikely event of a remote control malfunction. Please forward this information to all current and potential MEERA Operating Tables users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. " Whether or not you have affected product(s) with the Product REF numbers, please complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (800) 976-2664. Actions to be taken by Getinge: Maquet/Getinge ha

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.

Quantity

22 units