Respironics V60 Ventilator, manufactured by Respironics California Inc. in Carlsbad, CA.
Recall
- Recall Number
- Z-0457-2011
- Event Number
- 56147
- Firm
- Respironics California Inc
- FEI Number
- 2518422
- Product Code
- MNT
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 2, 2010
- Posted
- November 24, 2010
- Terminated
- December 1, 2010
- Address
- 2271 Cosmos Ct, Carlsbad, CA, 92011
Description
Respironics V60 Ventilator, manufactured by Respironics California Inc. in Carlsbad, CA.
Respironics California Inc has received customer complaints of an odor or burning smell coming from the V60 Ventilator, and identified a component on the Power Management PCB that exhibited thermal damage.
Phililps Healthcare issued an Urgent Medical Device Field Correction letter dated July 2, 2010 concerning the RespironicsV60 Ventilator issue. Customers will be contacted by US Field Service Technicians and international distributors to schedule and perform inspections and replacement of the affected component. The unit may continue to be used pending completion of the field inspection. Customers may contact Respironics US Customer Service about this action at 1 800 345 6443, option #5, then option #1.
Worldwide Distribution: Throughout the US and Puerto Rico,, and to the countries of: Australia, Austria, Belgium, Canada, Denmark, France, Germany, Hong Kong, India, Italy, Kenya, Kuwait, Libya, Malaysia, Oman, Netherlands, Norway, the United Kingdom, Saudi Arabia, and Spain.