FDA Recall Terminated

GE Medical Systems, COMPAX 40E X-ray Systems, Models; 2208203, 2208205, 2244282, 2244284, 2225863, 2225864 , 2225865, 2225870, 2225881, and 2225882. Intended to be used for general purpose radiography and tomography.

Recall: Z-0457-2010 · Initiated July 30, 2009

Recall

Recall Number
Z-0457-2010
Event Number
52720
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
IZZ
Status
Terminated
Root Cause
Other
Initiated
July 30, 2009
Posted
November 25, 2009
Terminated
May 19, 2016
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Medical Systems, COMPAX 40E X-ray Systems, Models; 2208203, 2208205, 2244282, 2244284, 2225863, 2225864 , 2225865, 2225870, 2225881, and 2225882. Intended to be used for general purpose radiography and tomography.

Reason

GE Healthcare has recently become aware that table lock safety features, which are provided with Compax 40E X-ray systems, are not always being utilized and has resulted in unintended lateral tabletop movement that may impact patient safety.

Action

Consignees in US and Canada were sent on July 31, 2009 a GE Healthcare "Urgent Medical device Correction" letter dated July 30, 2009. The letter was addressed to Hospital Administrators/ Risk mangers; Radiology Department Mangers; Radiologists. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product correction, and Contact Information. The letter instructs customers that the table manual mechanical lock is provided with the system and should be engaged when loading or unloading patients to ensure the tabletop is stabilized and not moving. For the rest of the world, the customer letter was dated October 22, 2009 and was mailed October 30, 2009.

Distribution

Worldwide Distribution -- USA, including states of AL, AK, AZ, AR, CA, CO, CT, DC,FL, GA, HI, ID, IL , IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH,NJ, NM, NY, NC, OH, OK OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WY and countries of UNITED KINGDOM, THAILAND, TAIWAN, SWITZERLAND, SPAIN, SAUDI ARABIA, PUERTO RICO, PORTUGAL NEW CALEDONIA, MARTINIQUE, KOREA, JAPAN, JAMAICA, ITALY, ISRAEL, HONG KONG, HONDURAS, GREECE, GERMANY, FRANCE, EGYPT, DOMINICAN, CHINA, CANDA, BRAZIL AUSTRIA, and AUSTRALIA.

Quantity

1046