i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.
Recall
- Recall Number
- Z-0456-2011
- Event Number
- 55178
- Firm
- Abbott Point of Care Inc.
- FEI Number
- 2245578
- Product Code
- CGL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 17, 2010
- Posted
- November 24, 2010
- Terminated
- September 22, 2011
- Address
- 400 College Rd E, Princeton, NJ, 08540-6607
Description
i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.
The i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks.
Abbott Point of Care issued Urgent Product Recall Notification letters dated March 2010 to all direct customers identifying the affected product and the actions to be taken. Customers were instructed to determine if they had cartirdges that had been stored above 25 degrees Celsius, and if so, contact Abbott for replacement or credit. Abbott Technical Support can be contacted at 800-366-8020 option 1 about this recall.
Worldwide Distribution: Throughout USA and Canada.
34525 cartridges