FDA Recall Terminated

Merit Medical Cultura Collection and Transport System, Ref CFTS100, Single Use, IVD, for the following unit counts: 1 Unit - (01)00884450492123, 50 Units - (01)10884450492120, 200 units - (01)20884450492127

Recall: Z-0455-2021 · Initiated September 16, 2020

Recall

Recall Number
Z-0455-2021
Event Number
86522
Firm
Merit Medical Systems, Inc.
FEI Number
1721504
Product Code
JSM
Status
Terminated
Root Cause
Device Design
Initiated
September 16, 2020
Terminated
June 1, 2021
Address
1600 W Merit Pkwy South, Jordan, UT, 84095-2416

Description

Merit Medical Cultura Collection and Transport System, Ref CFTS100, Single Use, IVD, for the following unit counts: 1 Unit - (01)00884450492123, 50 Units - (01)10884450492120, 200 units - (01)20884450492127

Reason

The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in treatment.

Action

The firm sent an recall notification on September 16, 2020 to all affected consignees. In addition to informing consignees about the correction, the firm requested consignees take the following action: 1. Please refer to the attached instructions and/or linked instructional video when collecting samples with nasopharyngeal swabs. 2. Ensure that applicable personnel within your organization are made aware of this notice. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. If you have any questions or concerns regarding this communication or the products addressed, please contact your Merit Sales Representative or Merit Customer Service at (801) 208-4381.

Distribution

Worldwide distribution. US Nationwide, Australia, Canada, Colombia, Ghana, Ireland, Jamaica, Kenya, Panama, Qatar, Trinidad and Tobago.

Quantity

159,396 units