FDA Recall Open, Classified

CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to add CARESCAPE ONE MBZ101 2087075-001

Recall: Z-0454-2026 · Initiated September 26, 2025

Recall

Recall Number
Z-0454-2026
Event Number
97786
Firm
GE Healthcare Finland Oy Kuortaneenkatu 2 Helsinki Finland
FEI Number
3002806783
Product Code
MHX
Status
Open, Classified
Root Cause
Component change control
Initiated
September 26, 2025
Posted
November 7, 2025

Description

CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to add CARESCAPE ONE MBZ101 2087075-001

Reason

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Action

On September 26, 2025, the recalling firm began notifying affected customers. A second letter was issued on October 3, 2025 which addressed a correction to Affected Product Details information. The customer letter advised users to use a monitor with a battery that is not within the affected battery population. However, pending correction by GE HealthCare, users may continue to use a monitor with affected batteries by following instructions included in the letter. These instructions include having mains power connected whenever possible, confirming appropriate battery status prior to use, and being aware that the monitor may still shut down during patient transport. Users are advised to be prepared to provide alternate methods of assessing the clinical status of the patient, bring a mains power cord during patient transport, and bring a spare battery. GE HealthCare will correct all affected products at no cost to the customer. A GE HealthCare representative will contact each customer to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide - US Nationwide distribution.

Quantity

5935