MeritMedical Cultura Flex Swab, Sterile Flocked Collection Device, Ref C100F, Sterile, for the following unit count:1 unit - (01)00884450491102, 25 unit - (01)10884450491109, 100 unit - (01)20884450491106
Recall
- Recall Number
- Z-0454-2021
- Event Number
- 86522
- Firm
- Merit Medical Systems, Inc.
- FEI Number
- 1721504
- Product Code
- KXG
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 16, 2020
- Terminated
- June 1, 2021
- Address
- 1600 W Merit Pkwy South, Jordan, UT, 84095-2416
Description
MeritMedical Cultura Flex Swab, Sterile Flocked Collection Device, Ref C100F, Sterile, for the following unit count:1 unit - (01)00884450491102, 25 unit - (01)10884450491109, 100 unit - (01)20884450491106
The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in treatment.
The firm sent an recall notification on September 16, 2020 to all affected consignees. In addition to informing consignees about the correction, the firm requested consignees take the following action: 1. Please refer to the attached instructions and/or linked instructional video when collecting samples with nasopharyngeal swabs. 2. Ensure that applicable personnel within your organization are made aware of this notice. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. If you have any questions or concerns regarding this communication or the products addressed, please contact your Merit Sales Representative or Merit Customer Service at (801) 208-4381.
Worldwide distribution. US Nationwide, Australia, Canada, Colombia, Ghana, Ireland, Jamaica, Kenya, Panama, Qatar, Trinidad and Tobago.
3,385,442 units