Freedom Boom models series: F110, F120, F130, F320, F330, F350, F410, and F420 with built in 42A Lightweight Spring Arm part numbers 42A-43L, 42A-43M, 42A-43H, and 42A-43S. Ceiling mounted utilities and equipment dispensing system.

Recall: Z-0452-2024 · Initiated July 19, 2023

Recall

Recall Number
Z-0452-2024
Event Number
93351
Firm
IHB OPERATIONS B.V. Floresstraat 52 Zwolle Netherlands
FEI Number
3006800085
Product Code
FQO
Status
Open, Classified
Root Cause
Device Design
Initiated
July 19, 2023
Posted
December 1, 2023

Description

Freedom Boom models series: F110, F120, F130, F320, F330, F350, F410, and F420 with built in 42A Lightweight Spring Arm part numbers 42A-43L, 42A-43M, 42A-43H, and 42A-43S. Ceiling mounted utilities and equipment dispensing system.

Reason

Weakened connection point on the adjustable suspension arm, potentially causing damage or a break.

Action

A recall notification issued October 09, 2023 via certified mail . Follow-up will be made by service representatives within 2 weeks to arrange for correction to replace all bearing assemblies in the field with an updated design. Prior to the field correction, inspect your Freedom 42A Lightweight Spring Arm for any movement in the end bearing assembly. If any loose connections are identified, please immediately remove the Spring Arm from use and notify your local authorized representative. Address questions or concerns to your local authorized representative or Skytron at 1-800-759-8766.

Distribution

US: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MB, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, VT, WA, WI, WV.

Quantity

1,497