BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Recall
- Recall Number
- Z-0451-2024
- Event Number
- 93360
- Firm
- BioFire Diagnostics, LLC
- FEI Number
- 3002773840
- Product Code
- PEN
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- October 17, 2023
- Posted
- December 1, 2023
- Address
- 515 S Colorow Dr, Salt Lake City, UT, 84108-1248
Description
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
On 10/17/23, "URGENT: MEDICAL DEVICE CORRECTION" notices were mailed to customers who were asked to do the following: 1) Review a complete list of BIOFIRE Quality Control Materials and references, for Example IQCP Risk Analysis can be found in BIOFIRE QC Tech Note (FLM1-PRT-0251), available from bioMrieux customer support. 2) Assess your Individualized Quality Control Plan (IQCP) and prepare for supplemental C. tropicalis testing if a Candida tropicalis Not Detected result is reported when testing with the MMQCI FilmArray BCID2 Control Panel M416 and/or amend testing plans to include alternate Quality Control materials such as those in FLM1-PRT-0251. 3) Complete and return the via email to [email protected] Contact customer support at [email protected] or via telephone by dialing 1.800.735.6544 and selecting option 5 for Product Technical Support with any questions or concerns.
Worldwide Distribution: US (nationwide): DE, OH, NY, IL, KY, FL, GA, KS, CA, WI, NC,, AK, PA, MN, ND, IN, AR, CO, UT, TX, AL, NJ, MD, SD, LA, OR, MA, ID, NE, MT, VA, IA, MO, WV, TN, MI, WY, SC, AZ, WA, NH, OK, HI, MS, DC, VT, CT, NM, RI, ME, NV; and OUS (foreign) to countries of: Algeria, Angola, Armenia, Bahrain, Bolivia, Bosnia- Herzegovina, Brunei, Bulgaria, Costa Rica, Croatia, Cyprus, Dominican Republic, Dutch Antilles, Ecuador, El Salvador, Estonia, Gabon, Georgia, Guam, Guatemala, Honduras, Indonesia, Ireland, Iceland, Israel, Ivory Coast, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macao, Macedonia, Malaysia, Mali, Mongolia, Morocco, New Zealand, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Qatar, Romania, Saudi Arabia, Serbia , Slovenia, Tunisia, Uganda, United Arab Emirate, Uruguay, Venezuela, Vietnam, Zimbabwe, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Spain, France, United Kingdom, Greece, India, Italia, Japan, Korea, Netherlands, Poland, Portugal, Sweden, Singapore, Thailand, Turkey, Taiwan, Switzerland, Austria, Denmark, Finland, Germany, Belgium, Hungary, South Africa, Philippines, Norway, Mexico
32,910