FDA Recall Terminated

Mobile X-ray system MobileDaRt Evolution/FDR Go Software The device is a mobile X-ray system and allows a technician to take a general digital radiography of patients in the hospital or medical facility who cannot move and outpatients in emergency.

Recall: Z-0451-2016 · Initiated November 10, 2015

Recall

Recall Number
Z-0451-2016
Event Number
72683
Firm
Shimadzu Medical Systems
FEI Number
1000125362
Product Code
IZL
Status
Terminated
Root Cause
Software design
Initiated
November 10, 2015
Terminated
October 27, 2016
Address
20101 S Vermont Ave, Torrance, CA, 90502-1328

Description

Mobile X-ray system MobileDaRt Evolution/FDR Go Software The device is a mobile X-ray system and allows a technician to take a general digital radiography of patients in the hospital or medical facility who cannot move and outpatients in emergency.

Reason

The firm is recalling the Shimadzu Mobile X-ray system because an image may not transfer to image server properly.

Action

Shimadzu sent an Urgent Voluntary Medical Device Recall Notice dated November 15, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter was sent to customers to inform them that Shimadzu is recalling the Mobile X-ray system MobileDaRt Evolution because of reports that an image was not transferred to image server properly. Customers with questions are instructed to contact their local Shimadzu representative or contact Shimadzu National Technical Support at (800) 228-1429.

Distribution

US Distribution and Internationally to Canada

Quantity

297 units