MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.
Recall
- Recall Number
- Z-0447-2016
- Event Number
- 72725
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 24, 2015
- Terminated
- September 15, 2021
- Address
- 400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227
Description
MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.
First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.
Elekta sent an Urgent Important Field Safety Notice 371-01-MSQ-004 to all affected customers on 11/24/2015. The notice informs users to the issue, the work around and the solution. The customers are instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta Office or Representative, as soon as possible but no longer than 30 days from receipt.
Nationwide Distribution -- AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, & Puerto Rico.