FDA Recall Open, Classified

Computed Tomography X-Ray System, Model uCT 530, Rx, CE, GTIN: 06971576831012 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Recall: Z-0445-2021 · Initiated August 5, 2020

Recall

Recall Number
Z-0445-2021
Event Number
86269
Firm
Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China
FEI Number
3011015597
Product Code
JAK
Status
Open, Classified
Root Cause
Software design
Initiated
August 5, 2020

Description

Computed Tomography X-Ray System, Model uCT 530, Rx, CE, GTIN: 06971576831012 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Reason

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Action

The firm sent notification letters via email to distributorsThe firm's consignee hand delivered a written Field Safety Notification to end-users on June 11, 2020 to inform them that the delay time displayed on the DDP can be incorrect for a timed scan such as a multi-phase contrast scan. The DDP is located on the gantry in the scan room. If the operator enters the scan room to perform any multi-phase contrast scan and refers to the incorrect delay time displayed on the DDP, he or she may not be able to leave the scan room before the scan starts, and as a result may receive unnecessary radiation dose. The letter states the following: The purpose of this letter is to advise you that United Imaging is voluntarily correcting some of the uCT 530, uCT 550, uCT 760 The uCT 530 and uMI 550 devices. Reason for the Voluntary Correction: We are aware of one complaint associated with the device. We have not received any report that have resulted in injury. Safety Issue: A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. Risk to Health of Safety Issue: The delay time displayed on the DDP can be incorrect for a timed scan such as a multi-phase contrast scan. The DDP is located on the gantry in the scan room. If the operator enters the scan room to perform any multi-phase contrast scan and refers to the incorrect delay time displayed on the DDP, he or she may not be able to leave the scan room before the scan starts, and as a result may receive unnecessary radiation dose. Actions to be taken by the Customer I User: 1. During a multi-phase contrast scan, do not perform any operations on the Preview image of any previous phases. 2. If possible, avoid entering the scan room when performing multi-phase contrast scans. 3. If it is necessary to enter the scan room when performing a multi-phase contrast scan, ignore the potentially incorrect delay time displayed on the

Distribution

US Nationwide distribution including in the state of TX.

Quantity

4