FDA Recall Open, Classified

Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (China); and Catalys Precision Laser System No Bed, Models: 0160-6020 (United States), 0160-6010 (International)

Recall: Z-0444-2022 · Initiated October 21, 2021

Recall

Recall Number
Z-0444-2022
Event Number
88986
Firm
AMO Manufacturing USA, LLC
FEI Number
3006695864
Product Code
OOE
Status
Open, Classified
Root Cause
Software design
Initiated
October 21, 2021
Address
510 Cottonwood Dr, Milpitas, CA, 95035-7403

Description

Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (China); and Catalys Precision Laser System No Bed, Models: 0160-6020 (United States), 0160-6010 (International)

Reason

Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.

Action

On 09/02/21, Microsoft "PrintNightmare" Vulnerability notices were sent to customers informing them of the upcoming software patches to remediate the vulnerability. This notice was posted on the recalling firm's website in the Security Advisories section. On 10/21/21 correction notices started being sent to customers. Customers were informed that Technical Service Representatives would contact then to schedule installation of the Windows OS patch. Customers were asked to do/ensure the following: 1) Complete the customer reply form, 2) To safeguard patients form potential harm and to inform them what to do in the event of product malfunction, customers were directed to specific parts of the Instructions for Use. 3) Confirm all measurements against baseline pre-operative manifest refraction and do not proceed when outside of recommended limits of difference, for sphere, cylinder and axis, as specified in the operators manual and training material. 4) Maintain manual records of patient measurements. 5) Ensure you comply with obligations under applicable privacy laws relating to putting in place safeguards in connection with personal information for patients. Customers with additional questions were encouraged to call Technical Support at 1 (800)- 511-0911.

Distribution

US: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM, CA. OUS: Canada, Mexico, Russian Federation, Turkey, Brazil, Egypt, Iraq, Saudi Arabia, Jordan, Norway, Colombia, Germany, Malaysia, India, Oman, Azerbaijan, Canada, Slovakia, Lebanon, Spain, Ireland, Italy, South Korea, Bangladesh, Chile, Taiwan, Tunisia, Kazakhstan, Libya, Argentina, Singapore, Austria, Australia, Vietnam, Greece, Poland, Indonesia, Yemen, Great Britain, South Africa, Croatia, France, Czech Republic, Algeria, Thailand, Bulgaria, United Arab Emirates, Belgium, Japan, China, Nepal, Israel, Switzerland, Dominican Republic, Hong Kong, Portugal, Ecuador, Kuwait, Netherlands, Slovenia, Philippines

Quantity

546