FDA Recall Open, Classified

Arrow AC3 Optimus Intra-Aortic Balloon Pump, AC3 Optimus IABP NA/AJLA, REF IAP-0701 (IPN917286), cardiac pump

Recall: Z-0443-2023 · Initiated October 17, 2022

Recall

Recall Number
Z-0443-2023
Event Number
91021
Firm
ARROW INTERNATIONAL Inc.
FEI Number
3013162291
Product Code
DSP
Status
Open, Classified
Root Cause
Pending
Initiated
October 17, 2022
Posted
December 15, 2022
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437

Description

Arrow AC3 Optimus Intra-Aortic Balloon Pump, AC3 Optimus IABP NA/AJLA, REF IAP-0701 (IPN917286), cardiac pump

Reason

There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.

Action

The firm issued an URGENT MEDICAL DEVICE NOTIFICATION to its consignees on 10/17/2022 by FedEx 2-day mail. To reduce the risk of short battery run time: .Plug IABP into an AC outlet during patient use, and when not in use. .Ensure the battery is fully charged prior to transporting patients, and a fully charged backup IABP device is available. .Replace the batteries when Battery Run Time is less than 90 minutes, there is visual damage to the battery, or Battery has been in service for 3 years. . Perform Battery Load Test at least every 12 months by qualified service personnel. If there is an issue with the battery load, immediately quarantine the device and contact Teleflex Customer Service to report the issue and receive service support for the affected IABP device. If a battery load test has not been performed in the past 12 months, perform battery load test prior to transporting patients with affected IABP devices. .If the IABP device battery fails while in use, immediately connect to an AC power source to continue therapy. .If a source of AC power is not available, transfer the patient to a fully charged back-up IABP device. .If pumping cannot be restored within 15 to 30 minutes, manually inflate and deflate the IAB several times per hour to reduce the risk of thrombus formation and consider removing the balloon. Facility Actions: Send a copy of this notification to all relevant personnel in your organization. Place a copy of this Notice with all affected products. Immediately check your inventory of Arrow AutoCAT2 and Arrow AC3 Optimus IABP, whether stored or in use. Return the Acknowledgement Form to Teleflex Customer Service. Hours of operation are 8am-7pm EST. Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected].

Distribution

Worldwide