FDA Recall Terminated

Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (TArt and TVen), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct).

Recall: Z-0443-2021 · Initiated September 30, 2020

Recall

Recall Number
Z-0443-2021
Event Number
86577
Firm
Maquet Cardiovascular Us Sales, Llc
FEI Number
3008355164
Product Code
DRY
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
September 30, 2020
Terminated
May 25, 2023
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (TArt and TVen), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct).

Reason

The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.

Action

On September 30, 2020, the firm issued an Urgent Medical Device Recall letter to all affected consignees. Customers were informed that the printed circuit board assembly for the affected devices are not in compliance with the IEC standard. Actions to be taken by the user: -Please examine your inventory immediately to determine if you have a BMU 40 with any of the affected serial numbers. The serial number tag is on the rear of the device. -Your facility may continue to use the BMU 40 pending replacement of the PCBA Connector, by adhering to the following recommendations: Use the equipotential bonding connector on the rear of the BMU 40 Properly ground any accessory equipment connect to the BMU 40 Do not connect external devices to the BMU 40 serial ports (Com 1 /Com 2) with an operational/ communication voltage exceeding 25 VDC. See Figure 2 below. Use standard PPE (e.g. nitrile gloves) while operating the BMU 40 Follow IFU recommendations including avoid simultaneous contact with the patient and device. Getinge has developed a correction that will be available in Q1 2021. Should you require a loaner unit while service is being performed on your affected unit, Getinge will provide you a loaner unit when available, prior to the return of your affected unit to Getinge. Please complete and sign the included MEDICAL DEVICE CORRECTION RESPONSE FORM Getinge has developed a correction that will be available in Q1 2021. Once available, Getinge will contact the person you listed on the Response Form to arrange shipment of the affected BMU 40 unit to the Getinge Service Center in New Jersey. The BMU 40 will be returned to your facility upon completion of the correction. You may also call Getinge Tech Support at (888) 9GETUSA / (888) 9438872 Option 4,2,1 between the hours of 8:00 a.m. and 6:00 p.m. Eastern Standard Time.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AZ, CA, IA, MD, NC, PA, TX, WA, WI and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Ecuador, Finland, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Italy, Jordan, Malaysia, Malta, Netherlands, Norway, Pakistan, Peru, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela.

Quantity

51 units