FDA Recall Open, Classified

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;

Recall: Z-0442-2026 · Initiated October 10, 2025

Recall

Recall Number
Z-0442-2026
Event Number
97882
FEI Number
1417592
Product Code
CAI
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
October 10, 2025
Posted
November 13, 2025
Address
3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753

Description

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;

Reason

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Action

Safety notice was provided to customers on 10/10/2025. Follow-up "URGENT MEDICAL DEVICE RECALL" notification dated 10/21/2025 was sent out on 10/21/2025. Notification was through email and first-class mail. Customers were instructed to Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product and destroy affected kits on hand. No products are to be returned to Medline. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-215-FG Recall Code: If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. An expanded "URGENT MEDICAL DEVICE RECALL" notification dated 10/24/2025 was provided to customers on 10/24/25, which included additional affected products. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Distribution

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Quantity

96,406 total units