FDA Recall Terminated

IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.

Recall: Z-0438-2013 · Initiated April 25, 2011

Recall

Recall Number
Z-0438-2013
Event Number
62918
Firm
Extremity Medical LLC
FEI Number
3007289093
Product Code
HTO
Status
Terminated
Root Cause
Employee error
Initiated
April 25, 2011
Posted
November 26, 2012
Terminated
March 21, 2013
Address
300 Interpace Pkwy, Ste 410A, Parsippany, NJ, 07054-1148

Description

IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.

Reason

Product Event Reports revealed a broken reamer experienced intra-operatively. Pre-drilling was not performed prior to reaming.

Action

Extremity Medical sent a notification by e-mail that describes the "Slight modification to the Surgical Technique" on April 25, 2011. For questions regarading this recall call 973-588-8980.

Distribution

Worldwide Distribution - USA including FL, MA, NC, CA, MD, VA, IN, AL, NJ, AZ, OK, MT, NY, OH, IA, MI, TX, WI, GA, IL, TN and NE. Internationally to Switzerland.

Quantity

156 units