IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.
Recall: Z-0438-2013
·
Initiated April 25, 2011
Recall
- Recall Number
- Z-0438-2013
- Event Number
- 62918
- Firm
- Extremity Medical LLC
- FEI Number
- 3007289093
- Product Code
- HTO
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- April 25, 2011
- Posted
- November 26, 2012
- Terminated
- March 21, 2013
- Address
- 300 Interpace Pkwy, Ste 410A, Parsippany, NJ, 07054-1148
Description
IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.
Reason
Product Event Reports revealed a broken reamer experienced intra-operatively. Pre-drilling was not performed prior to reaming.
Action
Extremity Medical sent a notification by e-mail that describes the "Slight modification to the Surgical Technique" on April 25, 2011. For questions regarading this recall call 973-588-8980.
Distribution
Worldwide Distribution - USA including FL, MA, NC, CA, MD, VA, IN, AL, NJ, AZ, OK, MT, NY, OH, IA, MI, TX, WI, GA, IL, TN and NE. Internationally to Switzerland.
Quantity
156 units