FDA Recall Open, Classified

UUltrasound System 2300, bkActiv w/battery; Model No. 2300-11; System, Imaging, Pulsed Doppler, Ultrasonic

Recall: Z-0437-2026 · Initiated October 9, 2025

Recall

Recall Number
Z-0437-2026
Event Number
97830
Firm
B-K Medical A/S Mileparken 34 Herlev Denmark
FEI Number
3003705156
Product Code
IYN
Status
Open, Classified
Root Cause
Device Design
Initiated
October 9, 2025
Posted
November 6, 2025

Description

UUltrasound System 2300, bkActiv w/battery; Model No. 2300-11; System, Imaging, Pulsed Doppler, Ultrasonic

Reason

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Action

GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION NOTICE" to its consignees on 10/09/2025 via traceable means. The notice explained the issue, potential risk, and requested: "Actions to be taken by Customer / User You can continue to use your device by following the instructions below: 1. Review and follow the applicable addendums to the Instructions for Use and the updated Service Manuals. The Addendums and Service Manuals are available for download at: https://www.gehealthcare.com/support/manuals. 2. Follow the conditions defined within the addendum and Service Manual applicable to your product for further instructions. 3. Place the applicable addendum and Service Manual with your existing product Instructions for Use. 4. If you have difficulty accessing the documentation or require a printed copy of the applicable addendum or Service Manual, please contact a GE HealthCare Service Representative for assistance. Please ensure all potential users in your facility are made aware of this safety notification and the required actions. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare at 1-800-645-4442 or [email protected] or your local Service Representative.

Distribution

Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkiye, Ukraine, United Arab Emirates, and United Kingdom.

Quantity

567 units