FDA Recall Terminated

Panorama Patient Monitoring Network, Cardiac Arrhythmia Monitor.

Recall: Z-0437-06 · Initiated December 7, 2005

Recall

Recall Number
Z-0437-06
Event Number
34320
Firm
Datascope Corp
FEI Number
2221819
Product Code
MHX
Status
Terminated
Root Cause
Other
Initiated
December 7, 2005
Posted
January 24, 2006
Terminated
August 1, 2006
Address
800 MacArthur Blvd, Mahwah, NJ, 07430-2001

Description

Panorama Patient Monitoring Network, Cardiac Arrhythmia Monitor.

Reason

Software anomaly that affects the Panorama Telepack where if communications are lost during standby, certain alarms which appear set could be actually off.

Action

Datascope sent letters to the distributors and end users on 12/7/2005. This communication states that a Datascope Service Representative will conduct and document the corrective action during a visit to the customer.

Distribution

The product was distributed to hospitals and clinics nationwide. There is one US Government account, VA Medical Center, Birmingham, AL.

Quantity

110 sites