Panorama Patient Monitoring Network, Cardiac Arrhythmia Monitor.
Recall: Z-0437-06
·
Initiated December 7, 2005
Recall
- Recall Number
- Z-0437-06
- Event Number
- 34320
- Firm
- Datascope Corp
- FEI Number
- 2221819
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 7, 2005
- Posted
- January 24, 2006
- Terminated
- August 1, 2006
- Address
- 800 MacArthur Blvd, Mahwah, NJ, 07430-2001
Description
Panorama Patient Monitoring Network, Cardiac Arrhythmia Monitor.
Reason
Software anomaly that affects the Panorama Telepack where if communications are lost during standby, certain alarms which appear set could be actually off.
Action
Datascope sent letters to the distributors and end users on 12/7/2005. This communication states that a Datascope Service Representative will conduct and document the corrective action during a visit to the customer.
Distribution
The product was distributed to hospitals and clinics nationwide. There is one US Government account, VA Medical Center, Birmingham, AL.
Quantity
110 sites