Smiths Pneupac babyPAC portable ventilator, model 100. Designed for use by qualified medical caregivers, paramedics and other trained personnel, for hospital, emergency and transport ventilation of patients during respiratory distress or insufficiency.
Recall
- Recall Number
- Z-0436-06
- Event Number
- 34275
- Firm
- Smiths Medical PM, Inc.
- FEI Number
- 2183502
- Product Code
- BTL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 5, 2005
- Posted
- January 24, 2006
- Terminated
- December 26, 2006
- Address
- N7w22025 Johnson Dr, Waukesha, WI, 53186-1856
Description
Smiths Pneupac babyPAC portable ventilator, model 100. Designed for use by qualified medical caregivers, paramedics and other trained personnel, for hospital, emergency and transport ventilation of patients during respiratory distress or insufficiency.
A potential safety related problem associated with the Pnuepac babyPac Ventilator model 100 has been discovered in that the interface between the air input hose and the air input fitting may leak, and if tightened too much, the air input fitting may loosen when trying to remove the air hose.
Consignees were provided a Safety Action Bulletin, number 05-SAB05, dated 12/05/05. This requests that affected devices be quarantined, Smiths Medical be contacted and devices ultimately be returned to Smiths Medical Pm, Inc. Service Departmentfor repair.
CA, CO, IN,MN, NY, PA, TN, TX