Smiths Pneupac babyPAC portable ventilator, model 100. Designed for use by qualified medical caregivers, paramedics and other trained personnel, for hospital, emergency and transport ventilation of patients during respiratory distress or insufficiency.

Recall: Z-0436-06 · Initiated December 5, 2005

Recall

Recall Number
Z-0436-06
Event Number
34275
Firm
Smiths Medical PM, Inc.
FEI Number
2183502
Product Code
BTL
Status
Terminated
Root Cause
Other
Initiated
December 5, 2005
Posted
January 24, 2006
Terminated
December 26, 2006
Address
N7w22025 Johnson Dr, Waukesha, WI, 53186-1856

Description

Smiths Pneupac babyPAC portable ventilator, model 100. Designed for use by qualified medical caregivers, paramedics and other trained personnel, for hospital, emergency and transport ventilation of patients during respiratory distress or insufficiency.

Reason

A potential safety related problem associated with the Pnuepac babyPac Ventilator model 100 has been discovered in that the interface between the air input hose and the air input fitting may leak, and if tightened too much, the air input fitting may loosen when trying to remove the air hose.

Action

Consignees were provided a Safety Action Bulletin, number 05-SAB05, dated 12/05/05. This requests that affected devices be quarantined, Smiths Medical be contacted and devices ultimately be returned to Smiths Medical Pm, Inc. Service Departmentfor repair.

Distribution

CA, CO, IN,MN, NY, PA, TN, TX

Quantity

44