FDA Recall Open, Classified

CEREPAK Detachable Coil System - Name/REF: UNIFORM 2MM X 2.5 CM/FCX100202, UNIFORM 2.5MM X 3.3 CM/FCX100253, UNIFORM 3MM X 5.4 CM/FCX100305, UNIFORM 3.5MM X 6.6 CM/FCX100356, UNIFORM 4MM X 7.5 CM/FCX100407, UNIFORM 4MM X 11.5 CM/FCX100412, UNIFORM 5MM X 9.7 CM/FCX100509, UNIFORM 5MM X 17 CM/FCX100517, UNIFORM 6MM X 11.9 CM/FCX100611, UNIFORM 6MM X 26 CM/FCX100626, UNIFORM 7MM X 13.9 CM/FCX100713, UNIFORM 7MM X 30 CM/FCX100730, UNIFORM 8MM X 16.1 CM/FCX100816, UNIFORM 8MM X 29 CM/FCX100829, UNIFORM 3D 2MM X 2.5 CM/FCX140225, UNIFORM 3D 3MM X 3 CM/FCX140303, UNIFORM 3D 3MM X 4 CM/FCX140304, UNIFORM 3D 3MM X 6 CM/FCX140306, UNIFORM 3D 4MM X 6 CM/FCX140406, UNIFORM 3D 4MM X 8 CM/FCX140408, UNIFORM 3D 4MM X 4.5 CM/FCX140445, UNIFORM 3D 5MM X 7 CM/FCX140507, UNIFORM 3D 5MM X 12 CM/FCX140512, UNIFORM 3D 6MM X 9 CM/FCX140609, UNIFORM 3D 6MM X 15 CM/FCX140615, UNIFORM 3D 7MM X 11 CM/FCX140711, UNIFORM 3D 7MM X 17 CM/FCX140717, UNIFORM 3D 8MM X 12 CM/FCX140812, UNIFORM 3D 8MM X 20 CM/FCX140820, UNIFORM 3D 9MM X 14 CM/FCX140914, UNIFORM 3D 9MM X 22 CM/FCX140922, UNIFORM 3D 10MM X 16 CM/FCX141016, UNIFORM 3D 10MM X 25 CM/FCX141025, UNIFORM 3D 11MM X 18 CM/FCX141118, UNIFORM 3D 11MM X 27 CM/FCX141127, UNIFORM 3D 12MM X 19 CM/FCX141219, UNIFORM 3D 12MM X 30 CM/FCX141230, UNIFORM XL 5MM X 10 CM/FCX180510, UNIFORM XL 6MM X 12.2 CM/FCX180612, UNIFORM XL 7MM X 14.3 CM/FCX180714, UNIFORM XL 8MM X 13.6 CM/FCX180813, UNIFORM XL 8MM X 30 CM/FCX180830, UNIFORM XL 9MM X 15.3 CM/FCX180915, UNIFORM XL 9MM X 33 CM/FCX180933, UNIFORM XL 10MM X 17.1 CM/FCX181017, UNIFORM XL 10MM X 34 CM/FCX181034, UNIFORM XL 11MM X 18.9 CM/FCX181118, UNIFORM XL 11MM X 37 CM/FCX181137, UNIFORM XL 12MM X 20.7 CM/FCX181220, UNIFORM XL 12MM X 40 CM/FCX181240, UNIFORM XL 13MM X 43 CM/FCX181343, UNIFORM XL 14MM X 47 CM/FCX181447, UNIFORM XL 15MM X 50 CM/FCX181550, UNIFORM XL 16MM X 47 CM/FCX181647, UNIFORM XL 17MM X 50 CM/FCX181750, UNIFORM XL 18MM X 46 CM/FCX181846, UNIFORM XL 19MM X 50 CM/FCX181950, UNIFORM XL 20MM X 50 CM/FCX182050, FREEFORM MINI 1MM X 1CM/MCR091010, FREEFORM MINI 1MM X 1.5CM/MCR091015, FREEFORM MINI 1MM X 2CM/MCR091020, FREEFORM MINI 1MM X 2.5CM/MCR091025, FREEFORM MINI 1MM X 3CM/MCR091030, FREEFORM MINI 1MM X 4CM/MCR091040, FREEFORM MINI 1.5MM X 2CM/MCR091520, FREEFORM MINI 1.5MM X 2.5CM/MCR091525, FREEFORM MINI 1.5MM X 3CM/MCR091530, FREEFORM MINI 1.5MM X 4CM/MCR091540, FREEFORM MINI 2MM X 2CM/MCR092020, FREEFORM MINI 2MM X 3CM/MCR092030, FREEFORM MINI 2MM X 4CM/MCR092040, FREEFORM MINI 2MM X 6CM/MCR092060, FREEFORM MINI 2.5MM X 3.5CM/MCR092535, FREEFORM MINI 2.5MM X 4.5CM/MCR092545, FREEFORM MINI 2.5MM X 5.5CM/MCR092555, FREEFORM MINI 3MM X 4CM/MCR093040, FREEFORM MINI 3MM X 6CM/MCR093060, FREEFORM MINI 3MM X 8CM/MCR093080, FREEFORM 2MM X 2 CM/SCR120202, FREEFORM 3MM X 6 CM/SCR120306, FREEFORM 3MM X 8 CM/SCR120308, FREEFORM 4MM X 7 CM/SCR120407, FREEFORM 4MM X 10 CM/SCR120410, FREEFORM 4MM X 12 CM/SCR120412, FREEFORM 5MM X 10 CM/SCR120510, FREEFORM 5MM X 15 CM/SCR120515, FREEFORM 6MM X 10 CM/SCR120610, FREEFORM 6MM X 15 CM/SCR120615, FREEFORM 6MM X 20 CM/SCR120620, FREEFORM 7MM X 15 CM/SCR120715, FREEFORM 7MM X 21 CM/SCR120721, FREEFORM 8MM X 15 CM/SCR120815, FREEFORM 8MM X 24 CM/SCR120824, FREEFORM 9MM X 25 CM/SCR120925, FREEFORM 10MM X 30 CM/SCR121030, FREEFORM 12MM X 30 CM/SCR121230, FREEFORM 2.5MM X 5 CM/SCR122505, FREEFORM 2.5MM X 3.5 CM/SCR122535, FREEFORM 3.5MM X 9 CM/SCR123509, FREEFORM 3.5MM X 7.5 CM/SCR123575, HELIFORM SOFT 1.5MM X 4 CM/SHD100154, HELIFORM SOFT 2MM X 8 CM/SHD100208, HELIFORM SOFT 2.5MM X 10 CM/SHD100250, HELIFORM SOFT 2.5MM X 8 CM/SHD100258, HELIFORM SOFT 3MM X 8 CM/SHD100308, HELIFORM SOFT 3MM X 10 CM/SHD100310, HELIFORM SOFT 4MM X 8 CM/SHD100408, HELIFORM SOFT 4MM X 10 CM/ SHD100410, HELIFORM SOFT 5MM X 10 CM/ SHD100510, HELIFORM SOFT 5MM X 15 CM/ SHD100515, HELIFORM SOFT 6MM X 12 CM/ SHD100612, HELIFORM SOFT 6MM X 16 CM/ SHD100616, HELIFORM SOFT 7MM X 16 CM/ SHD100716, HELIFORM SOFT 7MM X 20 CM/ SHD100720, HELIFORM SOFT 8MM

Recall: Z-0434-2026 · Initiated October 2, 2025

Recall

Recall Number
Z-0434-2026
Event Number
97814
Firm
Cerenovus Inc
FEI Number
3007628272
Product Code
HCG
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
October 2, 2025
Posted
November 13, 2025
Address
6303 Blue Lagoon Dr, Ste 315, Miami, FL, 33126-6068

Description

CEREPAK Detachable Coil System - Name/REF: UNIFORM 2MM X 2.5 CM/FCX100202, UNIFORM 2.5MM X 3.3 CM/FCX100253, UNIFORM 3MM X 5.4 CM/FCX100305, UNIFORM 3.5MM X 6.6 CM/FCX100356, UNIFORM 4MM X 7.5 CM/FCX100407, UNIFORM 4MM X 11.5 CM/FCX100412, UNIFORM 5MM X 9.7 CM/FCX100509, UNIFORM 5MM X 17 CM/FCX100517, UNIFORM 6MM X 11.9 CM/FCX100611, UNIFORM 6MM X 26 CM/FCX100626, UNIFORM 7MM X 13.9 CM/FCX100713, UNIFORM 7MM X 30 CM/FCX100730, UNIFORM 8MM X 16.1 CM/FCX100816, UNIFORM 8MM X 29 CM/FCX100829, UNIFORM 3D 2MM X 2.5 CM/FCX140225, UNIFORM 3D 3MM X 3 CM/FCX140303, UNIFORM 3D 3MM X 4 CM/FCX140304, UNIFORM 3D 3MM X 6 CM/FCX140306, UNIFORM 3D 4MM X 6 CM/FCX140406, UNIFORM 3D 4MM X 8 CM/FCX140408, UNIFORM 3D 4MM X 4.5 CM/FCX140445, UNIFORM 3D 5MM X 7 CM/FCX140507, UNIFORM 3D 5MM X 12 CM/FCX140512, UNIFORM 3D 6MM X 9 CM/FCX140609, UNIFORM 3D 6MM X 15 CM/FCX140615, UNIFORM 3D 7MM X 11 CM/FCX140711, UNIFORM 3D 7MM X 17 CM/FCX140717, UNIFORM 3D 8MM X 12 CM/FCX140812, UNIFORM 3D 8MM X 20 CM/FCX140820, UNIFORM 3D 9MM X 14 CM/FCX140914, UNIFORM 3D 9MM X 22 CM/FCX140922, UNIFORM 3D 10MM X 16 CM/FCX141016, UNIFORM 3D 10MM X 25 CM/FCX141025, UNIFORM 3D 11MM X 18 CM/FCX141118, UNIFORM 3D 11MM X 27 CM/FCX141127, UNIFORM 3D 12MM X 19 CM/FCX141219, UNIFORM 3D 12MM X 30 CM/FCX141230, UNIFORM XL 5MM X 10 CM/FCX180510, UNIFORM XL 6MM X 12.2 CM/FCX180612, UNIFORM XL 7MM X 14.3 CM/FCX180714, UNIFORM XL 8MM X 13.6 CM/FCX180813, UNIFORM XL 8MM X 30 CM/FCX180830, UNIFORM XL 9MM X 15.3 CM/FCX180915, UNIFORM XL 9MM X 33 CM/FCX180933, UNIFORM XL 10MM X 17.1 CM/FCX181017, UNIFORM XL 10MM X 34 CM/FCX181034, UNIFORM XL 11MM X 18.9 CM/FCX181118, UNIFORM XL 11MM X 37 CM/FCX181137, UNIFORM XL 12MM X 20.7 CM/FCX181220, UNIFORM XL 12MM X 40 CM/FCX181240, UNIFORM XL 13MM X 43 CM/FCX181343, UNIFORM XL 14MM X 47 CM/FCX181447, UNIFORM XL 15MM X 50 CM/FCX181550, UNIFORM XL 16MM X 47 CM/FCX181647, UNIFORM XL 17MM X 50 CM/FCX181750, UNIFORM XL 18MM X 46 CM/FCX181846, UNIFORM XL 19MM X 50 CM/FCX181950, UNIFORM XL 20MM X 50 CM/FCX182050, FREEFORM MINI 1MM X 1CM/MCR091010, FREEFORM MINI 1MM X 1.5CM/MCR091015, FREEFORM MINI 1MM X 2CM/MCR091020, FREEFORM MINI 1MM X 2.5CM/MCR091025, FREEFORM MINI 1MM X 3CM/MCR091030, FREEFORM MINI 1MM X 4CM/MCR091040, FREEFORM MINI 1.5MM X 2CM/MCR091520, FREEFORM MINI 1.5MM X 2.5CM/MCR091525, FREEFORM MINI 1.5MM X 3CM/MCR091530, FREEFORM MINI 1.5MM X 4CM/MCR091540, FREEFORM MINI 2MM X 2CM/MCR092020, FREEFORM MINI 2MM X 3CM/MCR092030, FREEFORM MINI 2MM X 4CM/MCR092040, FREEFORM MINI 2MM X 6CM/MCR092060, FREEFORM MINI 2.5MM X 3.5CM/MCR092535, FREEFORM MINI 2.5MM X 4.5CM/MCR092545, FREEFORM MINI 2.5MM X 5.5CM/MCR092555, FREEFORM MINI 3MM X 4CM/MCR093040, FREEFORM MINI 3MM X 6CM/MCR093060, FREEFORM MINI 3MM X 8CM/MCR093080, FREEFORM 2MM X 2 CM/SCR120202, FREEFORM 3MM X 6 CM/SCR120306, FREEFORM 3MM X 8 CM/SCR120308, FREEFORM 4MM X 7 CM/SCR120407, FREEFORM 4MM X 10 CM/SCR120410, FREEFORM 4MM X 12 CM/SCR120412, FREEFORM 5MM X 10 CM/SCR120510, FREEFORM 5MM X 15 CM/SCR120515, FREEFORM 6MM X 10 CM/SCR120610, FREEFORM 6MM X 15 CM/SCR120615, FREEFORM 6MM X 20 CM/SCR120620, FREEFORM 7MM X 15 CM/SCR120715, FREEFORM 7MM X 21 CM/SCR120721, FREEFORM 8MM X 15 CM/SCR120815, FREEFORM 8MM X 24 CM/SCR120824, FREEFORM 9MM X 25 CM/SCR120925, FREEFORM 10MM X 30 CM/SCR121030, FREEFORM 12MM X 30 CM/SCR121230, FREEFORM 2.5MM X 5 CM/SCR122505, FREEFORM 2.5MM X 3.5 CM/SCR122535, FREEFORM 3.5MM X 9 CM/SCR123509, FREEFORM 3.5MM X 7.5 CM/SCR123575, HELIFORM SOFT 1.5MM X 4 CM/SHD100154, HELIFORM SOFT 2MM X 8 CM/SHD100208, HELIFORM SOFT 2.5MM X 10 CM/SHD100250, HELIFORM SOFT 2.5MM X 8 CM/SHD100258, HELIFORM SOFT 3MM X 8 CM/SHD100308, HELIFORM SOFT 3MM X 10 CM/SHD100310, HELIFORM SOFT 4MM X 8 CM/SHD100408, HELIFORM SOFT 4MM X 10 CM/ SHD100410, HELIFORM SOFT 5MM X 10 CM/ SHD100510, HELIFORM SOFT 5MM X 15 CM/ SHD100515, HELIFORM SOFT 6MM X 12 CM/ SHD100612, HELIFORM SOFT 6MM X 16 CM/ SHD100616, HELIFORM SOFT 7MM X 16 CM/ SHD100716, HELIFORM SOFT 7MM X 20 CM/ SHD100720, HELIFORM SOFT 8MM

Reason

Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

Action

On 10/2/2025, Johnson & Johnson Med Tech mailed "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" notices dated 10/1/2025 to customers who were asked to do the following: 1. Immediately stop using the affected devices. 2. Identify and segregate all affected devices that are in inventory / consignment at your facility. 3. Firm sales representative will coordinate the removal of the product from your facility and will coordinate the credit, as applicable, for any customer-owned inventory. 4. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 5. If any product subject to this recall has been forwarded to another facility, contact that facility, provide a copy of this letter and notify J&J sales representative. 6. If the product has been successfully implanted, there is no additional action needed. 7. Complete and return the attached Business Reply Form in Attachment 2 to: [email protected]. Please return the BRF even if you do not have product subject to this recall. 8. Keep this notice visibly posted for awareness until all product subject to this recall have been returned. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. If you have additional questions regarding this medical device recall (removal), please contact firm at [email protected].

Distribution

US Nationwide distribution: CO, GA, MD, UT, OK, WI, MA, TN, SC, TX, CA, WV, LA, IL, NY, IA, AZ, AL, CT, WA, OH, ME, DC, FL, SD, KY, NJ, DE, NC, PA, MI, MO, VA, NH, NV and OR.

Quantity

12004