Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell Vantage" Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
Recall
- Recall Number
- Z-0433-2026
- Event Number
- 97705
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- HAW
- Status
- Open, Classified
- Root Cause
- Process change control
- Initiated
- September 30, 2025
- Posted
- November 5, 2025
- Address
- 400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227
Description
Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell Vantage" Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound
On 09/30/2025, the firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that specified serial numbers, there is an increased risk that paint flakes from the engravings may separate from the Leksell Vantage Arc System components. Separation of paint flakes is likely to occur during cleaning and sterilization. If paint flakes are not washed away during the cleaning and sterilization process, these could remain on the Leksell Vantage Arc System when used in surgery. This poses a risk of paint flakes entering the surgical area. This situation has not been observed in clinical practice, but the risk should be fully considered and mitigated.
Worldwide - U.S. Nationwide distribution in the states of AR, CA, CO, FL, ID, IL, LA, MA, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. The countries of Algeria, Australia, Austria, Belgium, China, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Republic of Korea, Kuwait, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
338