Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).
Recall
- Recall Number
- Z-0433-2007
- Event Number
- 36490
- Firm
- Terumo Cardiovascular Systems Corp
- FEI Number
- 1828100
- Product Code
- DTQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 16, 2006
- Posted
- February 14, 2007
- Terminated
- July 7, 2009
- Address
- 6200 Jackson Rd, Ann Arbor, MI, 48103-9586
Description
Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).
A 'battery needs service' message may be displayed, or the remaining battery time may incorrectly become zero, even when the battery is fully functional.
Consignees were sent a letter dated 11/10/06 informing them of the nature of the issues and that the firm would contact them to schedule a service call to install the software upgrade package. The software upgrade installation began at customers on 10/16/06, prior to issuance of the letter to consignees.
Worldwide distribution ---- including USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.