FDA Recall Terminated

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

Recall: Z-0433-2007 · Initiated October 16, 2006

Recall

Recall Number
Z-0433-2007
Event Number
36490
Firm
Terumo Cardiovascular Systems Corp
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Other
Initiated
October 16, 2006
Posted
February 14, 2007
Terminated
July 7, 2009
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586

Description

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

Reason

A 'battery needs service' message may be displayed, or the remaining battery time may incorrectly become zero, even when the battery is fully functional.

Action

Consignees were sent a letter dated 11/10/06 informing them of the nature of the issues and that the firm would contact them to schedule a service call to install the software upgrade package. The software upgrade installation began at customers on 10/16/06, prior to issuance of the letter to consignees.

Distribution

Worldwide distribution ---- including USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.

Quantity

207