FDA Recall Open, Classified

Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPolar Head, 43 mm OD, 22 mm Head; 3205.44.028 Klassic BiPolar Head, 44 mm OD, 28 mm Head; 3205.45.028 Klassic BiPolar Head, 45 mm OD, 28 mm Head; 3205.46.028 Klassic BiPolar Head, 46 mm OD, 28 mm Head; 3205.47.028 Klassic BiPolar Head, 47 mm OD, 28 mm Head; 3205.48.028 Klassic BiPolar Head, 48 mm OD, 28 mm Head; 3205.49.028 Klassic BiPolar Head, 49 mm OD, 28 mm Head; 3205.50.028 Klassic BiPolar Head, 50 mm OD, 28 mm Head; 3205.51.028 Klassic BiPolar Head, 51 mm OD, 28 mm Head; 3205.52.028 Klassic BiPolar Head, 52 mm OD, 28 mm Head; 3205.53.028 Klassic BiPolar Head, 53 mm OD, 28 mm Head; 3205.54.028 Klassic BiPolar Head, 54 mm OD, 28 mm Head; 3205.55.028 Klassic BiPolar Head, 55 mm OD, 28 mm Head; 3205.56.028 Klassic BiPolar Head, 56 mm OD, 28 mm Head; 3205.58.028 Klassic BiPolar Head, 58 mm OD, 28 mm Head; 3205.60.028 Klassic BiPolar Head, 60 mm OD, 28 mm Head

Recall: Z-0432-2026 · Initiated September 23, 2025

Recall

Recall Number
Z-0432-2026
Event Number
97708
Firm
Total Joint Orthopedics, Inc.
FEI Number
3008544874
Product Code
KWY
Status
Open, Classified
Root Cause
Packaging
Initiated
September 23, 2025
Posted
November 5, 2025
Address
1567 E Stratford Ave, Salt Lake City, UT, 84106-3559

Description

Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPolar Head, 43 mm OD, 22 mm Head; 3205.44.028 Klassic BiPolar Head, 44 mm OD, 28 mm Head; 3205.45.028 Klassic BiPolar Head, 45 mm OD, 28 mm Head; 3205.46.028 Klassic BiPolar Head, 46 mm OD, 28 mm Head; 3205.47.028 Klassic BiPolar Head, 47 mm OD, 28 mm Head; 3205.48.028 Klassic BiPolar Head, 48 mm OD, 28 mm Head; 3205.49.028 Klassic BiPolar Head, 49 mm OD, 28 mm Head; 3205.50.028 Klassic BiPolar Head, 50 mm OD, 28 mm Head; 3205.51.028 Klassic BiPolar Head, 51 mm OD, 28 mm Head; 3205.52.028 Klassic BiPolar Head, 52 mm OD, 28 mm Head; 3205.53.028 Klassic BiPolar Head, 53 mm OD, 28 mm Head; 3205.54.028 Klassic BiPolar Head, 54 mm OD, 28 mm Head; 3205.55.028 Klassic BiPolar Head, 55 mm OD, 28 mm Head; 3205.56.028 Klassic BiPolar Head, 56 mm OD, 28 mm Head; 3205.58.028 Klassic BiPolar Head, 58 mm OD, 28 mm Head; 3205.60.028 Klassic BiPolar Head, 60 mm OD, 28 mm Head

Reason

Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.

Action

On September 5, 2025 Total Joint issued customer a Field action notification. On September 23, 2025, Total Joint issued a Urgent Medical Device Recall Notification. Total Joint ask consignees to take the following actions: 1. Stop implanting the affected product. 2. Locate and quarantine all affected inventory. 3. With preliminary info we initially sent a Market Withdrawal notification on 9/05/2025 to initiate the quarantining and return of all BiPolar Heads in your possession. 4. A list of all BiPolar Heads in your possession as of 9/05/2025 is shown within the letter, if you have already returned them we appreciate your swift response, if not please return these items immediately. 5. Please complete and return the attached acknowledgment form confirming receipt of this notice and detailing the quantities returned. If inventory discrepancies are identified, please note that in the Return Response Box. 6. Please contact Total Joint Orthopedics at (801)486-6070 with any questions or concerns.

Distribution

US Nationwide distribution in the states of Ohio, Texas, Florida, Georgia, and North Carolina.

Quantity

103 units