FDA Recall Open, Classified

BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide

Recall: Z-0430-2026 · Initiated October 17, 2025

Recall

Recall Number
Z-0430-2026
Event Number
97669
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
FRN
Status
Open, Classified
Root Cause
Device Design
Initiated
October 17, 2025
Posted
November 20, 2025
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide

Reason

If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

Action

On 10/17/2025, correction notices were mailed and emailed: Distribute notice to clinical and non-clinical users. Safety Reminder: https://academy.bd.com/en-us/home. Provide notice to any facilities you may have sold or transferred devices to. Use only BD-authorized companies when renting infusion devices. Complete and return response form to [email protected] If dropped, biomedical engineering should perform testing as specified in Level of Testing Guidelines, including rate accuracy verification testing. If any test fails see troubleshooting section, see Technical Service Manual, Chapter 5 and 6. User Manuals: https://bd.com/self-service. Firm contacts: Technical Support 888-812-3229, [email protected] . Product Complaints: [email protected] BD Alaris Infusion System Safety Reminder: To ensure system remains in good operating condition, if a device or accessory is dropped or severely jarred, take the device out of use. Even if no visible damage is present, internal components may be affected. Return the device to Biomedical Engineering for inspection and repair. Visually inspect the system before each use. Check all visible surfaces and moving parts and if you observe damage or find the device does not function as expected, return it to biomedical engineering for repair. Do not use a device that appears to be damaged or does not function as expected. Return device to Biomedical Engineering for repair. Only biomedical engineering should perform annual maintenance as specified in the System Maintenance Software User Manual. Important in ensuring regular testing of any potentially dropped or damaged device that was not immediately removed from use to prevent infusion errors, protect patient safety, and promote proper function. All should understand the potential safety risks associated with using a device that may have sustained damage.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.

Quantity

2,574,396