Niobe ES System, Models 001-006000-1 and 001-006100-1

Recall: Z-0430-2021 · Initiated September 21, 2020

Recall

Recall Number
Z-0430-2021
Event Number
86555
Firm
Stereotaxis Inc
FEI Number
3003084417
Product Code
PJB
Status
Terminated
Root Cause
Device Design
Initiated
September 21, 2020
Terminated
November 13, 2023
Address
4320 Forest Park Ave, Ste 100, Saint Louis, MO, 63108-2979

Description

Niobe ES System, Models 001-006000-1 and 001-006100-1

Reason

There is a potential for fire in the inside computer cabinet.

Action

The firm will be contacting consignees via email and providing an electronic Urgent Field Safety Notice and Customer Reply Form. Consignees are asked for return the acknowledgement form to the firm within 30 days of receipt.

Distribution

US and OUS distribution: US - TN, MI, MN, KY, MO, IN, TX, NM, CA, NE, OH, NJ, CT, KS, OR, WA, WI, FL, IL, UT, PA, AZ, NY, GA, ID, OK, CO, LA OUS - EU, China, Australia, Singapore, Canada, South Africa, and Japan

Quantity

121