FDA Recall Terminated

0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride is used to prepare inoculum that is analyzed by the VITEK 2 instrument. The VITEK 2 user manual instructs all users to prepare inoculum with 0.45%-0.50% sodium chloride injections, and the system is designed to use 0.45%-0.50% sodium chloride concentration only.

Recall: Z-0430-2013 · Initiated October 17, 2012

Recall

Recall Number
Z-0430-2013
Event Number
63479
Firm
Biomerieux Inc
FEI Number
1950204
Status
Terminated
Root Cause
Employee error
Initiated
October 17, 2012
Posted
November 22, 2012
Terminated
December 5, 2013
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride is used to prepare inoculum that is analyzed by the VITEK 2 instrument. The VITEK 2 user manual instructs all users to prepare inoculum with 0.45%-0.50% sodium chloride injections, and the system is designed to use 0.45%-0.50% sodium chloride concentration only.

Reason

The firm inventoried and distributed 0.9% Sodium Chloride Saline bags as 0.45% Sodium Chloride Saline bags for use with the VITEK 2 system.

Action

Biomerieux sent an Urgent Product Removal Notice letter dated October 16, 2012 to all affected customers via certified mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the affected product and immediately discard any remaining inventory. The letter included instructions on receiving product replacement or credit. For questions or concerns contact your local BioMerieux representative.

Distribution

Worldwide Distribution - USA (nationwide) including the states of: AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, PA, RI, TN, TX, VA and WI. Distribution was also made to Canada, military facilities and other foriegn countries.

Quantity

199 cases/14-1000mL bags