FDA Recall Terminated

Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High). Minotrol 16 is a tri-level control intended for in vitro diagnostic use in monitoring the accuracy and precision on HORIBA Medical hematology blood cell counters using HORIBA Medical reagents. Minotrol 16 is an in vitro diagnostic reagent composed of human erythrocytes, mammalian leukocytes, and platelets in a plasma-like fluid with preservatives. Minotrol 16 is a stable material that provides a means of determining the accuracy and precision of hematology blood cell counters. It is handled in the same manner as a patient specimen. The assay tables are determined on validated instruments using the appropriate reagents

Recall: Z-0429-2013 · Initiated September 14, 2012

Recall

Recall Number
Z-0429-2013
Event Number
63539
Firm
Horiba Instruments, Inc dba Horiba Medical
FEI Number
3002698356
Product Code
JPK
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 14, 2012
Posted
November 21, 2012
Terminated
April 22, 2013
Address
34 Bunsen, Irvine, CA, 92618-4210

Description

Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High). Minotrol 16 is a tri-level control intended for in vitro diagnostic use in monitoring the accuracy and precision on HORIBA Medical hematology blood cell counters using HORIBA Medical reagents. Minotrol 16 is an in vitro diagnostic reagent composed of human erythrocytes, mammalian leukocytes, and platelets in a plasma-like fluid with preservatives. Minotrol 16 is a stable material that provides a means of determining the accuracy and precision of hematology blood cell counters. It is handled in the same manner as a patient specimen. The assay tables are determined on validated instruments using the appropriate reagents

Reason

HORIBA Medical Irvine Technical Support received reports from some customers that have experienced high recover values for Hemoglobin when using Minotrol 16 lot #MX090 (Level High). Customer communication was sent 10/01/2010.

Action

HORIBA Medical sent a Product Information Notice letter dated November 5, 2010 via Fax. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to follow the instructions included with the Product Information Notice. For questions customers were instructed to call the Technical Support Hotline at 1-888-903-5001. For questions regarding this recall call 949-453-0500, ext 208.

Distribution

Nationwide Distribution

Quantity

22,954