FDA Recall Terminated

MYLA MASTER DVD V4.0 CLI Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).

Recall: Z-0428-2016 · Initiated November 3, 2015

Recall

Recall Number
Z-0428-2016
Event Number
72643
Firm
bioMerieux, Inc.
FEI Number
3002769706
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
November 3, 2015
Terminated
December 4, 2017
Address
100 Rodolphe St, Durham, NC, 27712-9402

Description

MYLA MASTER DVD V4.0 CLI Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).

Reason

MYLA software connected to a Laboratory Information System (LIS) and a VITEK 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.

Action

bioMerieux sent an Urgent Product Correction Notice letter dated November10, 2015 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local bioMrieux Clinical Customer Service organization at (800) 682-2666.

Distribution

Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and countries of: Foreign accounts: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, Germany, Greece, Hungary, India, Italy, Japan, Korea, Netherlands, Poland, Portugal, Russia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, Angola, Puerto Rico, Republic of Belarus, Republic of Belarus, Benin, Bosnia, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Chad, Democratic Republic of Congo, Costa Rica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Gabon, Guam, Guatemala, Guatemala, Guinea, Honduras, Indonesia, Iraq, Israel, Ivory Coast, Kazakhstan, Kenya, Kenya, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malaysia, Mali, Portugal, Myanmar, Managua, Niger, Nigeria, Palestine, Philippines, Philippines, Qatar, Kingdom of Saudi Arabia, Senegal, Serbia, Slovakia, Togo, Tunisia, Ukraine, UAE, Vietnam, Vietnam, Ivory Coast and Zimbabwe.

Quantity

626 systems in total