Level 1 Normothermic I.V. Fluid Administration Sets, Product Code DI-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjunction with Level 1 Fast Flow Fluid Warmers (H1000, H-1025, H-1028, and H-1200), for the warming and delivery of blood and I.V. solutions, at flow rates of 75ml/min - 530 ml/min.
Recall
- Recall Number
- Z-0428-2012
- Event Number
- 60516
- Firm
- Smiths Medical ASD, Inc.
- FEI Number
- 1221261
- Product Code
- LGZ
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 21, 2011
- Posted
- January 11, 2012
- Terminated
- August 7, 2014
- Address
- 160 Weymouth St, Rockland, MA, 02370-1136
Description
Level 1 Normothermic I.V. Fluid Administration Sets, Product Code DI-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjunction with Level 1 Fast Flow Fluid Warmers (H1000, H-1025, H-1028, and H-1200), for the warming and delivery of blood and I.V. solutions, at flow rates of 75ml/min - 530 ml/min.
Disconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Smiths Medical notified consignees by letter dated November 28, 2011 titled: Urgent Field Safety Notice. The letter identified the affected product, the problem, and the actions to be taken. Customers were instructed to inspect inventory and segregate unused affected product. Distributors were instructed to immediately stop distributing and quarantine all inventory. The letters included a Confirmation Form to be complete and return, by fax to 781-610-9859 or by email to [email protected]. For questions consignees were instructed to contact Smiths Medical at 1781-763-9330.
Worldwide Distribution (USA) (nationwide) - Argentina, Austria, Australia, Belgium, Bahrain, Bermuda, Canada, Chile, China, Columbia, Czech Republic, Germany, Denmark, Estonia, Egypt, Finland, France, Great Britain, Guernsey, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Ireland, Israel India, Iceland, Italy, Japan, South Korea, Kazakhstan, Lebanon, Luxembourg, Mexico, Malaysia, New Caledonia, Netherlands, Norway, New Zealand, Panama, Poland, Portugal, Qatar, Singapore, Sweden, Slovak Republic, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, and Vatican City State.
7631 units