Custom Ultrasonics System 83 Plus 2 and Plus 9 Endoscope Washer/Disinfector.

Recall: Z-0426-2014 · Initiated March 12, 2013

Recall

Recall Number
Z-0426-2014
Event Number
65985
Firm
Custom Ultrasonics, Inc.
FEI Number
2523209
Product Code
KOG
Status
Terminated
Root Cause
Software design
Initiated
March 12, 2013
Posted
November 27, 2013
Terminated
June 26, 2014
Address
144 Railroad Dr, Ivyland, PA, 18974-1449

Description

Custom Ultrasonics System 83 Plus 2 and Plus 9 Endoscope Washer/Disinfector.

Reason

Certain System 83 Plus 2 and System 83 Plus 9 units running the Windows 7 operating system were not fully validated prior to distribution.

Action

A notification letter, dated February 26, 2013, was sent to direct accounts to inform them of the recall and that technicians would visit facilities to replace the Windows 7 operating system with an appropriately validated Windows XP operating system.

Distribution

Worldwide Distribution: USA (nationwide) and Internationally to: Canada, Slovakia Republic, Sydney and Peru.

Quantity

118