Signa Ovation, 0.35T MR Systems (MFO 1-4). Indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the musculoskeletal, vascular, cardiac, and neuro systems.
Recall
- Recall Number
- Z-0426-2011
- Event Number
- 56380
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- August 23, 2010
- Posted
- November 19, 2010
- Terminated
- July 13, 2012
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Signa Ovation, 0.35T MR Systems (MFO 1-4). Indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the musculoskeletal, vascular, cardiac, and neuro systems.
If the lateral table adjustment crank handle breaks off, it has sufficient attractive force to the magnet and may result in patient injury.
Consignees were sent on 8/23/10 a GE "Urgent Medical Device Correction" letter dated August 20, 2010. The letter was addressed to Hospital Administrators / Risk Mangers, Radiology Department Managers and Radiologists. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction, and Contact Information. Customers were asked to discontinue use of their MR system and to contact their local service representative, if the lateral table adjustment crank handle on the table becomes loose.
Worldwide Distribution -- USA, including states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, KS, KY, LA, MD, MA, MI, MN, MS,MO,MT, NV, NH, NJ, NY, NC, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WV, and WI and countries of VENEZUELA, ECUADOR, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, SPAIN, RUSSIAN FEDERATION, REPUBLIC OF KOREA, POLAND, PHILIPPINE,S OMAN, MEXICO, KUWAIT, JAPAN, ITALY, HUNGARY, GERMANY, EGYPT, DENMARK, CHILE, CHINA, and BRASIL.
202