FDA Recall Terminated

Signa Ovation, 0.35T MR Systems (MFO 1-4). Indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the musculoskeletal, vascular, cardiac, and neuro systems.

Recall: Z-0426-2011 · Initiated August 23, 2010

Recall

Recall Number
Z-0426-2011
Event Number
56380
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
LNH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 23, 2010
Posted
November 19, 2010
Terminated
July 13, 2012
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Signa Ovation, 0.35T MR Systems (MFO 1-4). Indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the musculoskeletal, vascular, cardiac, and neuro systems.

Reason

If the lateral table adjustment crank handle breaks off, it has sufficient attractive force to the magnet and may result in patient injury.

Action

Consignees were sent on 8/23/10 a GE "Urgent Medical Device Correction" letter dated August 20, 2010. The letter was addressed to Hospital Administrators / Risk Mangers, Radiology Department Managers and Radiologists. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction, and Contact Information. Customers were asked to discontinue use of their MR system and to contact their local service representative, if the lateral table adjustment crank handle on the table becomes loose.

Distribution

Worldwide Distribution -- USA, including states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, KS, KY, LA, MD, MA, MI, MN, MS,MO,MT, NV, NH, NJ, NY, NC, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WV, and WI and countries of VENEZUELA, ECUADOR, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, SPAIN, RUSSIAN FEDERATION, REPUBLIC OF KOREA, POLAND, PHILIPPINE,S OMAN, MEXICO, KUWAIT, JAPAN, ITALY, HUNGARY, GERMANY, EGYPT, DENMARK, CHILE, CHINA, and BRASIL.

Quantity

202