MYLA CLINIC PATCH 3.2.0 CD Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).
Recall
- Recall Number
- Z-0425-2016
- Event Number
- 72643
- Firm
- bioMerieux, Inc.
- FEI Number
- 3002769706
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 3, 2015
- Terminated
- December 4, 2017
- Address
- 100 Rodolphe St, Durham, NC, 27712-9402
Description
MYLA CLINIC PATCH 3.2.0 CD Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).
MYLA software connected to a Laboratory Information System (LIS) and a VITEK 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.
bioMerieux sent an Urgent Product Correction Notice letter dated November10, 2015 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local bioMrieux Clinical Customer Service organization at (800) 682-2666.
Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and countries of: Foreign accounts: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, Germany, Greece, Hungary, India, Italy, Japan, Korea, Netherlands, Poland, Portugal, Russia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, Angola, Puerto Rico, Republic of Belarus, Republic of Belarus, Benin, Bosnia, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Chad, Democratic Republic of Congo, Costa Rica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Gabon, Guam, Guatemala, Guatemala, Guinea, Honduras, Indonesia, Iraq, Israel, Ivory Coast, Kazakhstan, Kenya, Kenya, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malaysia, Mali, Portugal, Myanmar, Managua, Niger, Nigeria, Palestine, Philippines, Philippines, Qatar, Kingdom of Saudi Arabia, Senegal, Serbia, Slovakia, Togo, Tunisia, Ukraine, UAE, Vietnam, Vietnam, Ivory Coast and Zimbabwe.
626 systems in total