FDA Recall Terminated

20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Recall: Z-0424-2014 · Initiated April 4, 2013

Recall

Recall Number
Z-0424-2014
Event Number
65356
Firm
B. Braun Medical, Inc.
FEI Number
2523676
Product Code
LHI
Status
Terminated
Root Cause
Equipment maintenance
Initiated
April 4, 2013
Posted
November 27, 2013
Terminated
August 26, 2014
Address
901 Marcon Boulevard, Allentown, PA, 18109-9512

Description

20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Reason

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

Action

B Braun sent an Urgent - Medical Device Removal Notification letter dated April 4, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by he customer. Customers were instructed to review the Product Recall Notification in its entirety and ensure that all users in their organization are informed about this voluntary product recall. Determine their current inventory of the affected lots within their distribution center. Contact their customers that have received the affected lots and inform them of this voluntary product recall. Combine both their inventory, and the inventory reported by their customers, and utilize their specific "Product Removal Acknowledgement" form to record the total number of individual units (within partial cases) and the number of full-unopened cases. If no inventory remaining enter zero (0). Return the completed Acknowledgement Form to BBMI Quality Assurance in the self addressed envelope or fax to 610-849-1197. Customers should call Customer Support Department at 800-227-2862 to arrange for eturn and replacement product or for questions. For questions regarding this recall call 800-227-2862.

Distribution

Nationwide Distribution

Quantity

1,746,079 units