FDA Recall Terminated

AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03

Recall: Z-0421-03 · Initiated December 18, 2002

Recall

Recall Number
Z-0421-03
Event Number
25055
Firm
Abbott Laboratories, Inc
FEI Number
1628664
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
December 18, 2002
Posted
January 11, 2003
Terminated
April 7, 2004
Address
1921 Hurd Drive PO Box 2020, Irving, TX, 75038

Description

AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03

Reason

Gas Springs supporting processing center cover lose pressure after repeated usage causing the cover to unexpectedly close.

Action

Abbott Laboratories sent a December 18, 2002 letter to all consignees informing consignees of their plans to install a locking mechanism on all AxSYM Systems. The installations will be performed by Abbott Representatives. A customer reply form was included in the device correction letter.

Distribution

Nationwide and to the following foreign countries: Mexico, Guatemala, Venezuela, Columbia, Ecuador, Peru, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Canada, Germany, Ireland, England, Belgium, France, Austria, Italy, Spain, Russia, Japan, Singapore, Hong Kong, Taiwan, Thailand, Malaysia, Korea, Australia, New Zeland, Honduras, Costa Rica, Trinidad, Bahamas, Northern Antilles, Bermuda, Panama, West Indies, Paraguay, Norway, Kingstow, VC

Quantity

18,832 units