AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03
Recall
- Recall Number
- Z-0421-03
- Event Number
- 25055
- Firm
- Abbott Laboratories, Inc
- FEI Number
- 1628664
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 18, 2002
- Posted
- January 11, 2003
- Terminated
- April 7, 2004
- Address
- 1921 Hurd Drive PO Box 2020, Irving, TX, 75038
Description
AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03
Gas Springs supporting processing center cover lose pressure after repeated usage causing the cover to unexpectedly close.
Abbott Laboratories sent a December 18, 2002 letter to all consignees informing consignees of their plans to install a locking mechanism on all AxSYM Systems. The installations will be performed by Abbott Representatives. A customer reply form was included in the device correction letter.
Nationwide and to the following foreign countries: Mexico, Guatemala, Venezuela, Columbia, Ecuador, Peru, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Canada, Germany, Ireland, England, Belgium, France, Austria, Italy, Spain, Russia, Japan, Singapore, Hong Kong, Taiwan, Thailand, Malaysia, Korea, Australia, New Zeland, Honduras, Costa Rica, Trinidad, Bahamas, Northern Antilles, Bermuda, Panama, West Indies, Paraguay, Norway, Kingstow, VC
18,832 units