FDA Recall Terminated

LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5mm x 50mm; i. 6.5mm x 55mm; j. 7.5mm x 35mm; k. 7.5mm x 40mm; l. 7.5mm x 45mm; m. 7.5mm x 50mm; n. 7.5mm x 55mm; o. 8.5mm x 35mm; p. 8.5mm x 40mm; q. 8.5mm x 45mm; r. 8.5mm x 50mm; s.8.5mm x 55mm; t. 9.5mm x 35mm; u. 9.5mm x 40mm; v. 9.5mm x 45mm; w. 9.5mm x 50mm; x. 9.5mm x 55mm; Used in spinal fixation surgery.

Recall: Z-0420-2023 · Initiated January 27, 2022

Recall

Recall Number
Z-0420-2023
Event Number
91070
Firm
Integrity Implants Inc.
FEI Number
3012797630
Product Code
NKB
Status
Terminated
Root Cause
Device Design
Initiated
January 27, 2022
Posted
December 5, 2022
Terminated
October 23, 2023
Address
354 Hiatt Drive, Palm Beach Gardens, FL, 33418

Description

LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5mm x 50mm; i. 6.5mm x 55mm; j. 7.5mm x 35mm; k. 7.5mm x 40mm; l. 7.5mm x 45mm; m. 7.5mm x 50mm; n. 7.5mm x 55mm; o. 8.5mm x 35mm; p. 8.5mm x 40mm; q. 8.5mm x 45mm; r. 8.5mm x 50mm; s.8.5mm x 55mm; t. 9.5mm x 35mm; u. 9.5mm x 40mm; v. 9.5mm x 45mm; w. 9.5mm x 50mm; x. 9.5mm x 55mm; Used in spinal fixation surgery.

Reason

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Action

Integrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.

Distribution

Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.

Quantity

1,369 screws