FDA Recall Terminated

CXP Analysis Software, Versions 2.0 & 2.1: Part #722396 CXP Analysis Software Kit V2.1, Single User; Part #722395 CXP Analysis Software Kit V2.1, Single User Upgrade; Part #722397 CXP Analysis Software Kit V2.1, 3 User Pack; Part #722398 CXP Analysis Software Kit V2.1, 5 User Pack; Part #722399 CXP Analysis Software Kit V2. 1, 15 User Pack; Part # 722593 CXP Analysis Software Kit V2.1, 5 Network User Pack Part #722594 CXP Analysis Software Kit V2.1, 10 Network User Pack; Part #722595 CXP Analysis Software Kit V2.1, 15 Network User Pack; Part #629640 CXP Analysis Software Kit V2.0, Single User; Part #629639 CXP Analysis Software Kit V2.0, Single User Upgrade; Part #629641 CXP Analysis Software Kit V2.0, 3 User Pack; Part #629642 CXP Analysis Software Kit V2.0, 5 User Pack; Part #629643 CXP Analysis Software Kit V2.0, 15 User Pack; Part #629644 CXP Analysis Software Kit V2.0, 5 Network User Pack; Part #629645 CXP Analysis Software Kit V2.0, 10 Network User Pack; Part #629646 CXP Analysis Software Kit V2.0, 15 Network User Pack. The Cytomics FC 500 is a system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles.

Recall: Z-0420-2007 · Initiated August 29, 2006

Recall

Recall Number
Z-0420-2007
Event Number
37083
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Other
Initiated
August 29, 2006
Posted
February 8, 2007
Terminated
April 19, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

CXP Analysis Software, Versions 2.0 & 2.1: Part #722396 CXP Analysis Software Kit V2.1, Single User; Part #722395 CXP Analysis Software Kit V2.1, Single User Upgrade; Part #722397 CXP Analysis Software Kit V2.1, 3 User Pack; Part #722398 CXP Analysis Software Kit V2.1, 5 User Pack; Part #722399 CXP Analysis Software Kit V2. 1, 15 User Pack; Part # 722593 CXP Analysis Software Kit V2.1, 5 Network User Pack Part #722594 CXP Analysis Software Kit V2.1, 10 Network User Pack; Part #722595 CXP Analysis Software Kit V2.1, 15 Network User Pack; Part #629640 CXP Analysis Software Kit V2.0, Single User; Part #629639 CXP Analysis Software Kit V2.0, Single User Upgrade; Part #629641 CXP Analysis Software Kit V2.0, 3 User Pack; Part #629642 CXP Analysis Software Kit V2.0, 5 User Pack; Part #629643 CXP Analysis Software Kit V2.0, 15 User Pack; Part #629644 CXP Analysis Software Kit V2.0, 5 Network User Pack; Part #629645 CXP Analysis Software Kit V2.0, 10 Network User Pack; Part #629646 CXP Analysis Software Kit V2.0, 15 Network User Pack. The Cytomics FC 500 is a system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles.

Reason

Two workflow scenarios associated with renaming regions may produce incorrect results.

Action

A Product Corrective Action (PCA) letter was sent on the week of August 29, 2006 to all accounts that have an FC500 or have ordered CXP or MXP software to inform them that a) incorrect statistics are generated when a polygonal region with vertical re-entrant vertices is renamed using the Region Properties dialog box. b)when a region is copied from one plot to another, and then the new region is renamed, the statistics are not updated with the new region name. (Letter was sent by US mail). Letter includes immediate user preventive action to remedy both issues. A response form is included for consignees to return to recalling firm.

Distribution

Nationwide & Canada

Quantity

200 units nationwide