Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890
Recall
- Recall Number
- Z-0417-2007
- Event Number
- 37065
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- DDC
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 26, 2006
- Posted
- February 7, 2007
- Terminated
- April 30, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890
Low end imprecision affecting both S0 calibrators and patient samples which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay.
A Product Corrective Action (PCA) letter was mailed on Oct 27, 2006 to all Thyroglobulin Antibody customers. They are informed that there has been an increase in customer feedback regarding low end imprecision affecting both S0 calibrators and patient samples, which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay. The letter informs the customer that the firm recommends that they use an alternate method for testing TgAb. (Letter was sent by US mail). Customers are asked to complete and return the enclosed response form within 10 days.
Nationwide and Canada
474 units