FDA Recall Terminated

Guardia Access Nano Embryo Transfer Catheter is use for transferring IVF embryos into the uterine cavity.

Recall: Z-0415-2020 · Initiated October 3, 2019

Recall

Recall Number
Z-0415-2020
Event Number
84108
Firm
Cook Inc.
FEI Number
1820334
Product Code
MQF
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
October 3, 2019
Terminated
May 13, 2020
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

Guardia Access Nano Embryo Transfer Catheter is use for transferring IVF embryos into the uterine cavity.

Reason

Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.

Action

An Urgent Field Safety Notice letter dated 10/3/19 was sent to customers. The letter instructs customers to identify affected devices, quarantine the devices, return them, and complete the enclosed Customer Response Form. Where product is indicated as being returned, our Customer Services department will contact you to organize the return and issue you with the relevant Returns Authorization number. Please include contact details on the Customer Response form. Returned Product should be addressed to: Cook Medical EUDC Robert-Koch-Strae, 2 52499 Baesweiler GERMANY Credit will be provided for the returned affected products where applicable.

Distribution

No US distribution. The products were distributed to the following foreign countries: Belgium, Denmark, Germany, Italy, Netherlands, Norway, Poland, Republic of Macedonia, Slovenia, Spain, and Switzerland.

Quantity

202