Rascal LiteWay 214 manufactured by Giant Manufacturing, Taiwan.
Recall
- Recall Number
- Z-0415-2011
- Event Number
- 56978
- Firm
- Electric Mobility Corp
- FEI Number
- 2246975
- Product Code
- INI
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 27, 2010
- Posted
- November 18, 2010
- Terminated
- April 3, 2014
- Address
- 599 Mantua Boulevard, One Mobility Plaza, Sewell, NJ, 08080
Description
Rascal LiteWay 214 manufactured by Giant Manufacturing, Taiwan.
Scooter seat bases were found to have insufficient welding which may lead to the seat base becoming loose and causing the seat to wobble. This may result in complete failure of the seat base, causing the user to fall off the vehicle.
The Rascal Company (Electric Mobility Corporation) sent Urgent Medical Device Recall Notice Scooter Seat Base letters dated October 7, 2010 and November 11, 2010 letter to their affected retail customers and affected dealers. The letter instructs the reader that the seat base may become loose, causing the seat to wobble and if not replaced the wobble could lead to a complete failure of the seat base and cause an individual to fall off their vehicle and become injured. The letters also instruct the reader to replace seat bases immediately. Respone forms are attached. Seats with new seat bases will be shipped with return shipping labels to return the original base to the factory. Service representatives will replace the seat base for customers. Customers can contact the Rascal Company about this recall at 1-800-662-4548.
Worldwide Distribution: Throughout the US and Canada
172 units