Philips IntelliVue Clinical Information Portfolio (ICIP), Clinical Management System Software Version D.00 through D.03.
Recall
- Recall Number
- Z-0414-2011
- Event Number
- 57039
- Firm
- Philips Healthcare Inc.
- FEI Number
- 1218950
- Product Code
- DXJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- October 19, 2010
- Posted
- November 18, 2010
- Terminated
- August 1, 2017
- Address
- 3000 Minuteman Road, Andover, MA, 01810
Description
Philips IntelliVue Clinical Information Portfolio (ICIP), Clinical Management System Software Version D.00 through D.03.
Software: Scheduling function of the Philips Intellivue Clinical Information Portfolio Releases D.00 through D.03. Edits to a scheduled even frequency order may result in pending interventions not being generated after the first 24 hours.
Philips Healthcare issued an Urgent Medical Device Correction letter dated October 15, 2010 to customers. Customers are asked to follow the guidelines in the "Action to be taken by Customer/User" section of the Urgent Medical Device Correction Notice until they receive their software correction. The correction will be in the form of a software upgrade and will be performed by Philips service personnel. Contact Philips Customer Care Center with questions or concerns about this correction at 1-800-722-9377
Worldwide Distribution: USA, including states of CA, DC, FL, IL, IN, KY, MA, MD, ME, NJ, NY, OK, RI, TX, VA, VT, and WV; and the countries of: AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ITALY, KENYA, LATVIA, MALAYSIA, NETHERLANDS, NORWAY, SAUDI ARABIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, UNITED ARAB EMIRATES, and the UNITED KINGDOM.
200 units