FDA Recall Terminated

AutoSuture Bladed Trocar, equipped with safety shield. Model number 179094. Available with smooth or threaded sleeve in sizes 5-15mm inner diameter and 70-100mm length.

Recall: Z-0414-2007 · Initiated October 19, 2006

Recall

Recall Number
Z-0414-2007
Event Number
37162
Firm
Ascent Healthcare Solutions, Inc.
FEI Number
1000132435
Product Code
NLM
Status
Terminated
Root Cause
Process design
Initiated
October 19, 2006
Posted
February 8, 2007
Terminated
October 31, 2008
Address
5307 Great Oak Dr, Lakeland, FL, 33815

Description

AutoSuture Bladed Trocar, equipped with safety shield. Model number 179094. Available with smooth or threaded sleeve in sizes 5-15mm inner diameter and 70-100mm length.

Reason

The tip of the trocars may separate before and while in use. In some instances the trocars have been found with exposed blades upon opening the packaging.

Action

The recall was initiated on 10/19/2006 by direct distribution of the recall notice to direct accounts by the firm's sales force. The firm also posted the notification on their web site. All customers were instructed to examine their inventories for the product and to remove and quarantine all affected product. Instructions on returning the product to the firm were also included along with a response form.

Distribution

Nationwide

Quantity

1566